SUMMARY:Invited talk: A statistical reconstruction algorithm for positroni
 um lifetime imaging using time-of-flight positron emission tomography
DTSTART;VALUE=DATE-TIME:20220711T122000Z
DTEND;VALUE=DATE-TIME:20220711T124000Z
DTSTAMP;VALUE=DATE-TIME:20261008T195223Z
UID:indico-contribution-41-658@indico.koza.if.uj.edu.pl
DESCRIPTION:Speakers: Hsin-Hsiung Bill  Huang\; University of Central Flor
 ida\, USA ()\nAs an emerging technology\, positronium lifetime imaging (PL
 I) using time-of-flight (TOF) positron emission tomography (PET) provides 
 supplemental information reflecting the microenvironment surrounding the t
 issue\, in addition to the radiotracer activities from PET scan. However\,
  due to the finite coincidence resolving time (CRT)\, typically on the ord
 er of 200 – 600 ps\, reconstruction of the lifetime image from PLI list-
 mode events has yet to be fully developed. \n  Previous studies have estab
 lished the feasibility of positronium lifetime imaging using TOF PET scann
 ers that have superior CRT to allow precise localization of each event in 
 space [Moskal 2019\, Shibuya 2020]. Currently\, no practical TOF PET scann
 ers have the needed CRT and therefore the studies consider only well-separ
 ated sources that can be resolved by systems having a finite CRT. The exis
 ting positronium imaging method [Qi 2022] considers the marginal likelihoo
 d of the lifetime parameter given the radiotracer activity parameter and a
 pplies a surrogate function to estimate the lifetime parameter. In contras
 t\, our proposed method directly models the lifetime and activity paramete
 rs and estimates them jointly. Our method overcomes all these challenges a
 nd produces quantitatively correct lifetime images beyond the resolution l
 imit of the system CRT.\n  We address the problem of reconstructing the li
 fetime image from TOF PET data with finite CRT. We present a statistical r
 econstruction framework that maximizes the likelihood of observed PLI even
 ts\, each comprising the following observable attributes:  the detector re
 ceiving the prompt gamma\, the detector pair receiving the annihilation pa
 ir\, the time difference between the detection of prompt gamma and annihil
 ation pair\, and the time difference between the two photons within the an
 nihilation.  In our study\, our simulation data include the flight time of
  the photons before detection and our method does not require having sub-1
 00 ps TOF resolution for accurate localization of each detected event.  We
  have verified our method using PLI events generated by the Monte-Carlo si
 mulation.\n\nhttps://indico.koza.if.uj.edu.pl/event/7/contributions/658/
LOCATION:Collegium Maius
URL:https://indico.koza.if.uj.edu.pl/event/7/contributions/658/
END:VEVENT
BEGIN:VEVENT
SUMMARY:Current progress\, challenges and frontiers for Ps-based oncology 
 nanodiagnostics
DTSTART;VALUE=DATE-TIME:20220711T134000Z
DTEND;VALUE=DATE-TIME:20220711T140000Z
DTSTAMP;VALUE=DATE-TIME:20261008T195223Z
UID:indico-contribution-41-660@indico.koza.if.uj.edu.pl
DESCRIPTION:Speakers: Bożena Zgardzińska\; Maria Curie-Sklodowska Univer
 sity\, Poland ()\nPublication agreement (CC BY 4.0) 	Yes\nPrimary authors\
 nBożena Zgardzińska (Institute of Physics\, Maria Curie-Sklodowska Unive
 rsity) Dr Jerzy Mielko (Department of Surgical Oncology\, Medical Universi
 ty of Lublin) Ms Katarzyna Gęca (Department of Surgical Oncology\, Medica
 l University of Lublin) Marek Gorgol (Maria Curie-Sklodowska University\, 
 Institute of Physics\, Department of Nuclear Methods\, Pl. M. Curie-Sklodo
 wskiej 1\, 20-031 Lublin\, Poland) Radek Zaleski (Maria Curie-Skłodowska 
 University) Magdalena Goździuk (Maria Curie- Sklodowska University) Boże
 na Jasińska (Institute of Physics\, Maria Curie Sklodowska University) Dr
  Stephenson Owusu (Radiation Research Laboratory\, University of Notre Dam
 e) Dr Gustaw Chołubek (Diagnostic Techniques Unit\, Faculty of Health Sci
 ences\, Medical University of Lublin) Mr Konrad Wysogląd (Institute of Ph
 ysics\, Maria Curie-Sklodowska University) Katsiaryna Yankova\nPresentatio
 n Materials\nThere are no materials yet.\n\nhttps://indico.koza.if.uj.edu.
 pl/event/7/contributions/660/
LOCATION:Collegium Maius
URL:https://indico.koza.if.uj.edu.pl/event/7/contributions/660/
END:VEVENT
BEGIN:VEVENT
SUMMARY:Study of positronium in normal and cancer cells
DTSTART;VALUE=DATE-TIME:20220711T132000Z
DTEND;VALUE=DATE-TIME:20220711T134000Z
DTSTAMP;VALUE=DATE-TIME:20261008T195223Z
UID:indico-contribution-41-659@indico.koza.if.uj.edu.pl
DESCRIPTION:Speakers: Ewelina Kubicz-Staszkiewicz\; Jagiellonian Universit
 y\, Poland ()\nhttps://indico.koza.if.uj.edu.pl/event/7/contributions/659/
LOCATION:Collegium Maius
URL:https://indico.koza.if.uj.edu.pl/event/7/contributions/659/
END:VEVENT
BEGIN:VEVENT
SUMMARY:Multiphoton time-of-flight MLEM reconstruction for the positronium
  imaging in J-PET
DTSTART;VALUE=DATE-TIME:20220711T130000Z
DTEND;VALUE=DATE-TIME:20220711T132000Z
DTSTAMP;VALUE=DATE-TIME:20261008T195223Z
UID:indico-contribution-41-657@indico.koza.if.uj.edu.pl
DESCRIPTION:Speakers: Roman Shopa\; National Centre for Nuclear Research\,
  Poland ()\nWe modify the positronium imaging method for the Jagiellonian 
 PET (J-PET) scanners\, using the simulated data for the decays of a metast
 able ortho-positronium (o-Ps) based on the pick-off and conversion process
 es resulting in two 511-keV annihilation photons [1\, 2]. An analytical mo
 del\, embedded into the time-of-flight maximum likelihood expectation maxi
 misation (TOF MLEM) reconstruction algorithm\, is introduced for the asses
 sment of the multiphoton detector response and the estimation of the o-Ps 
 mean lifetime.\nThe detection probabilities of the J-PET system matrix are
  calculated for the coincident events each constituted by two annihilation
  and one deexcitation photons. TOF information is used for the iterative M
 LEM run and the estimation of a time difference between the annihilation a
 nd the positron emission\, the distribution of which is built for each vox
 el. The latter is updated in event-by-event way using the weights for the 
 ultimate MLEM iteration.\nEmploying the Geant4 software [3]\, we simulated
  the experiments with four point-like sources put into two existing J-PET 
 prototypes\, each corresponding to a tissue of a different o-Ps lifetime\,
  with total activity 1.1 MBq for the injected 22Na [2\, 4]. The decay mode
 l was restricted to include only the decays into back-to-back photons\, im
 plying the exponentially modified Gaussian shape for the aforementioned ti
 me difference histograms.\nUtilising TOF MLEM allowed to reduce the noise 
 and significantly improve the spatial resolution of the positronium imagin
 g\, compared to the earlier studies [1\, 2]. The o-Ps mean lifetimes\, cal
 culated as a result of fitting the histograms exhibited a good consistency
  with the simulation setup. The reconstruction algorithm uses the advanced
  resolution model that accounts for both the geometrical factors and the d
 etector blur of the J-PET\, and can as well be extended to include attenua
 tion correction or other contributions\, as well as for the o-Ps decay int
 o three photons.\n\n[1] P. Moskal et al. “Performance assessment of the 
 2γ positronium imaging with the total-body PET scanners\,” EJNMMI Phys.
 \, vol. 7\, no. 44\, pp. 1\, Jun. 2020.\n[2] P. Moskal et al.\, “Positro
 nium imaging with the novel multiphoton PET scanner\,” Sci. Adv.\, vol. 
 7\, no. 42\, pp. 1\, Oct. 2021.\n[3] S. Agostinelli et al.\, “Geant4 –
  A Simulation Toolkit\,” Nucl. Instrum. Meth. A\, vol. 506\, no. 1\, pp.
  250\, Jul. 2003.\n[4] P. Moskal et al.\, “Simulating NEMA characteristi
 cs of the modular total-body J-PET scanner – an economic total-body PET 
 from plastic scintillators\,” Phys. Med. Biol.\, vol. 66\, no. 17\, pp. 
 175015\, Sep. 2021.\n\nhttps://indico.koza.if.uj.edu.pl/event/7/contributi
 ons/657/
LOCATION:Collegium Maius
URL:https://indico.koza.if.uj.edu.pl/event/7/contributions/657/
END:VEVENT
BEGIN:VEVENT
SUMMARY:Invited talk: Positron Annihilation Spectroscopy of oxygen content
  tissue-equivalent samples
DTSTART;VALUE=DATE-TIME:20220711T124000Z
DTEND;VALUE=DATE-TIME:20220711T130000Z
DTSTAMP;VALUE=DATE-TIME:20261008T195223Z
UID:indico-contribution-41-582@indico.koza.if.uj.edu.pl
DESCRIPTION:Speakers: Ali Biganeh\; Nuclear Science and Technology Researc
 h Institute\, Iran ()\nPositron Annihilation Spectroscopy (PAS) has numero
 us applications in material science such as studies of structural defects\
 , free volumes\, porous materials\, and their chemical environments. In re
 cent years\, the applications of this technique in chemical processes\, ra
 diobiology\, and medicine have been the subject of research by many resear
 chers [1\, 2]. PAS includes Coincidence Doppler Broadening Spectroscopy (C
 DBS) of annihilation radiation to determine the momentum of the electrons 
 involved in the annihilation process and Positron Annihilation Lifetime Sp
 ectroscopy (PALS) for the investigation of the mechanism behind the positr
 onium annihilation in the sample. In this paper\, the possible application
  of the PALS technique for the characterization of healthy and cancer tiss
 ues is evaluated. The purpose of the experimental part of this work is to 
 investigate the sensitivity of the PALS technique to the oxygen concentrat
 ion in tissue-equivalent samples.\nThe accepted hypothesis for this resear
 ch is that the concentration of oxygen in cancer tissues is less than in h
 ealthy tissues due to the massive cell proliferation [3]. To conduct the e
 xperiments\, the PALS spectrometer based on digital signal processing has 
 been designed\, commissioned\, and optimized in the NSTRI positron lab. Th
 is spectrometer makes it possible to record the time difference of events 
 with a channel width of 10 ps and the timing resolution of 173 ps using th
 e plastic scintillation detectors. To determine the sensitivity of the tec
 hnique to the presence of the polar groups including carbon-oxygen boundin
 g\, 4 polymer samples were measured using PALS and CDBS techniques\, and t
 he correlation between the parameters was clarified. The results of this i
 nvestigation confirmed that the position and height of the peak in the Orb
 ital Electron Momentum Spectrum (OEMS) obtained in the Doppler experiment 
 and the component of the positronium lifetime in the PALS technique are bo
 th sensitive to the presence of oxygen in the content [4]. To determine th
 e PALS parameters related to the amount of oxygen\, 6 tissue-equivalent po
 lymers with different concentrations of oxygen were selected. The positron
  lifetime spectra in these samples were measured using the developed digit
 al PALS spectrometer and the results were analyzed using LT-10 software. T
 he results of this experiment confirmed that in general\, as the oxygen co
 ncentration in the sample increases\, due to the quenching role of oxygen 
 atoms in the formation of the positronium atoms\, the I_3 and τ_3 paramet
 ers decrease significantly. However\, in the samples with a difference of 
 about 4% in the oxygen concentration\, the behavior analysis of the positr
 onium lifetime and its intensity requires a systematic study of the effect
  of density\, porosity\, crystallinity\, and the bond type in the polymer 
 samples. The results of this work apply to the development of a tumor imag
 ing system based on the PALS technique.\nReferences\n[1] Moskal\, Paweł\,
  and Ewa Stępień. "Positronium as a biomarker of hypoxia." Bio-Algorithm
 s and Med-Systems (2021).‏\n[2] Stepanov\, P. S.\, et al. "Interaction o
 f positronium with dissolved oxygen in liquids." Physical Chemistry Chemic
 al Physics 22.9 (2020): 5123-5131.‏\n[3] Brahimi-Horn\, M. Christiane\, 
 Johanna Chiche\, and Jacques Pouysségur. "Hypoxia and cancer." Journal of
  molecular medicine 85.12 (2007): 1301-1307.‏\n[4] Zare\, Mahshid\, et a
 l. "Experimental investigation of the oxygen sensing ability of positron a
 nnihilation spectroscopy for tumor imaging." Radiation Physics and Enginee
 ring (2022).‏\n\nhttps://indico.koza.if.uj.edu.pl/event/7/contributions/
 582/
LOCATION:Collegium Maius
URL:https://indico.koza.if.uj.edu.pl/event/7/contributions/582/
END:VEVENT
BEGIN:VEVENT
SUMMARY:First clinical positronium imaging of patients
DTSTART;VALUE=DATE-TIME:20220711T120000Z
DTEND;VALUE=DATE-TIME:20220711T122000Z
DTSTAMP;VALUE=DATE-TIME:20261008T195223Z
UID:indico-contribution-41-581@indico.koza.if.uj.edu.pl
DESCRIPTION:Speakers: Pawel Moskal\; Jagiellonian University\, Poland ()\n
 https://indico.koza.if.uj.edu.pl/event/7/contributions/581/
LOCATION:Collegium Maius
URL:https://indico.koza.if.uj.edu.pl/event/7/contributions/581/
END:VEVENT
END:VCALENDAR